PM Procedure - Wall-Mounted Vacuum Unit (General)

Physical, Mechanical, Electrical, and Functionality Tests
Med-Stat
05/08/2023

PM Procedure - Wall-Mounted Vacuum Unit (General)

Physical, Mechanical, Electrical, and Functionality Tests

    Control Number

    PM PROCEDURE

    Inspect exterior of unit for damage/corrosion/cleanliness, clean as needed

    • Pass
    • Flag
    • Fail

    Clean tubing by passing cold disinfection solution through suction canister

    • Pass
    • Flag
    • Fail

    Turn regulator knob and test correlating needle response

    • Pass
    • Flag
    • Fail

    Test regulator at 60mm/hg with calibrated external vacuum gauge (+/-5%)

    • Pass
    • Flag
    • Fail

    Test regulator at 140mm/hg with calibrated external vacuum gauge (+/-5%)

    • Pass
    • Flag
    • Fail

    Test regulator at 200mm/hg with calibrated external vacuum gauge (+/-5%)

    • Pass
    • Flag
    • Fail

    Test regulator at 180mm/hg on "INT" setting; timing module engages between 6-10s; needle moves to and remains at 180mmHg for 14-18s

    • Pass
    • Flag
    • Fail

    Additional Notes

Source: Med-Stat (Community Member)

Please note that any procedure, checklist, or other document available in the MaintainX Procedure Hub is provided for general education and information only and does not constitute legal, medical, or financial advice. MaintainX makes the materials available AS IS and AS PROVIDED, without warranties of any kind. By downloading or using any such materials, you assume the risk that they may not be appropriate for your specific situation and agree that you are solely responsible for any such use, including compliance with applicable law and with meeting any conditions of product warranties.