Verification Record of Absolute Viscometer
Procedure ACMP-OP-007: Equipment and apparatus used in testing materials must be calibrated or verified for accuracy on a regular basis to assure the equipment is giving reliable results.
Verification Record of Absolute Viscometer
Procedure ACMP-OP-007: Equipment and apparatus used in testing materials must be calibrated or verified for accuracy on a regular basis to assure the equipment is giving reliable results.
Date
Performed By
Manufacturer
Lab ID
Model
Thermometer ID
Viscometer Bath
Bath filled with clean distilled water or clear oil
Maintains temperature of 60 +/- 0.1 degree C
Actual Temp
Thermometer correctly immersed at test temperature
Vacuum System
Maintains vacuum of 300 +/- 0.5 mmHG
Actual mmHg
Holds vacuum when system is evacuated and closed
Standardized at least once per year
Thermometer
ASTM 47C or 47F used
Other thermometric device with an accuracy of at least 0.02 degree C used
Thermometer or other device calibrated at least every six months
Viscometers
Clean and free of deposits or contaminates
Immersed so top timing mark at least 20 mm below surface of bath liquid at test temperature
Cleaned with suitable solvent and properly dried or cleaned in a glass cleaning oven not set above 500 degree C
Verified with proper viscosity standard at least every five years
Date of last verification
Comments