Ductilometer Calibration Record
A calibration record for ductilometers, documenting inspection, verification, and maintenance procedures, including results and recommended actions.
Ductilometer Calibration Record
A calibration record for ductilometers, documenting inspection, verification, and maintenance procedures, including results and recommended actions.
Verified By
Date
Equipment/Model
Serial No.
Verification Equipment Used
Results
Visually inspect the ductility bath for cleanliness and level of water
Reset the sample port dial indicator to zero
Run the machine until the indicator reaches 50 cm and record time
Verify rate of travel per min. (5 cm +/- 5%)
Verify bath water can be maintained within +/- 0.5 degrees C of set point
Maintenance Items
Clean bath and replace water if necessary
Place a couple of drops of lubricating oil in oil ports
Action Recommended
Comments
Source: Ergon - Lubbock 14842 (Community Member)