Verification Record of DSR
Use "Verifying Dynamic Shear Rheometer (Anton Paar)" procedure. Attach verification records for GMH 3710 and wafer, along with DSR temperature calibration report and viscosity standard reports to this record.
Verification Record of DSR
Use "Verifying Dynamic Shear Rheometer (Anton Paar)" procedure. Attach verification records for GMH 3710 and wafer, along with DSR temperature calibration report and viscosity standard reports to this record.
Verified By
Date
Verification Equipment Used (SN)
Caliper
GMH 3710
Wafer
Cannon Certified Viscosity Standard N2700000SP
Plate Diameter 25mm Bottom
Tolerance 25 +/- 0.05mm
S/N:
Measurement 1
Measurement 2
Measurement 3
Plate Diameter 25mm Top
Tolerance 25 +/- 0.05mm
S/N:
Measurement 1
Measuremant 2
Measurement 3
Plate Diameter 8mm Bottom
Tolerance 8 +/- 0.02mm
S/N:
Measurement 1
Measurement 2
Measurement 3
Plate Diameter 8mm Top
Tolerance 8 +/- 0.02mm
S/N:
Measurement 1
Measurement 2
Measurement 3
Plates
Upper and lower plates aligned
Upper plate does not wobble
Temperature Calibration
Run temperature calibration as instructed by manufacturer: Temperature range between 4 and 88 degrees C
Standard value @ 64 C as noted on bottle
ODSR Cannon Standard @ 64 C
Standard within +/- 3% ?
File print outs
Maintenance
Counter cooling system flushed
Coiled fins cleaned
Air supply dry and filtered
DSR protected from electrical surges
Action Recommened
Comments
Source: Ergon - Lubbock 14842 (Community Member)