Back

Verification Record of DSR

Use "Verifying Dynamic Shear Rheometer (Anton Paar)" procedure. Attach verification records for GMH 3710 and wafer, along with DSR temperature calibration report and viscosity standard reports to this record.
Ergon - Lubbock 14842
06/19/2024

Verification Record of DSR

Use "Verifying Dynamic Shear Rheometer (Anton Paar)" procedure. Attach verification records for GMH 3710 and wafer, along with DSR temperature calibration report and viscosity standard reports to this record.

    Verified By

    Date

    Verification Equipment Used (SN)

    Caliper

    GMH 3710

    Wafer

    Cannon Certified Viscosity Standard N2700000SP

    Plate Diameter 25mm Bottom

    Tolerance 25 +/- 0.05mm

    S/N:

    Measurement 1

    Measurement 2

    Measurement 3

    Plate Diameter 25mm Top

    Tolerance 25 +/- 0.05mm

    S/N:

    Measurement 1

    Measuremant 2

    Measurement 3

    Plate Diameter 8mm Bottom

    Tolerance 8 +/- 0.02mm

    S/N:

    Measurement 1

    Measurement 2

    Measurement 3

    Plate Diameter 8mm Top

    Tolerance 8 +/- 0.02mm

    S/N:

    Measurement 1

    Measurement 2

    Measurement 3

    Plates

    Upper and lower plates aligned

    • Yes
    • No
    • N / A

    Upper plate does not wobble

    • Yes
    • No
    • N / A

    Temperature Calibration

    Run temperature calibration as instructed by manufacturer: Temperature range between 4 and 88 degrees C

    Standard value @ 64 C as noted on bottle

    ODSR Cannon Standard @ 64 C

    Standard within +/- 3% ?

    • Pass
    • Flag
    • Fail

    File print outs

    • Yes
    • No
    • N / A

    Maintenance

    Counter cooling system flushed

    • Yes
    • No
    • N / A

    Coiled fins cleaned

    • Yes
    • No
    • N / A

    Air supply dry and filtered

    • Yes
    • No
    • N / A

    DSR protected from electrical surges

    • Yes
    • No
    • N / A

    Action Recommened

    Comments

Source: Ergon - Lubbock 14842 (Community Member)

Please note that any procedure, checklist, or other document available in the MaintainX Procedure Hub is provided for general education and information only and does not constitute legal, medical, or financial advice. MaintainX makes the materials available AS IS and AS PROVIDED, without warranties of any kind. By downloading or using any such materials, you assume the risk that they may not be appropriate for your specific situation and agree that you are solely responsible for any such use, including compliance with applicable law and with meeting any conditions of product warranties.